
![]()
|
|
| GeoFast FID (Version 2.12) |
| Calibration and Setup Procedure |
| ITEC GEO0086 |
| Page 1/17 |
|
|
| FSD – Field Support Department bruary 2009 05 Fe |
|
|
| Issued b |
|
|
| Subject: The GeoFastFID (version 2.12) calibration procedure is somewhat similar to the |
| Reserval II procedure. The Calibration port of the GFF gas panel cannot be used to |
| calibrate the chromatograph five points calibration instead the manual injection port is |
| used. A new quick look procedure to do the Chromatograph and Total Gas |
| calibrations is documented here. |
|
|
| Operating System: The GFF version 2.12 can operate under Windows 2000 (611.AA.760) or Windows |
| XP (700.BV.519) operating system environment. |
|
|
| Port Setting: The communication ports and speed settings are very important both on the GFF |
| computer and also on the RTM computer. |
|
|
| 1. On the GFF computer – COM 1 port setting: |
|
|
|
|

|
|
| Go to ‘Control Panel’, double click on ‘System’ |
| yVerified by Approved by |
| FSD R&T Dept OSL Manager |
| Revision 0 |
| 05 Feb 2009 20 Feb 2009 02 Mar 2009 |
| ITEC GEO 0086L14 200T |
|
|
|
|
|
|

|
|
| Click on ‘Hardware’ |

![]()
|
|
| GeoFast FID (Version 2.12) |
| Calibration and Setup Procedure |
| ITEC GEO0086 |
| Page 2/17 |
|
|
| FSD – Field Support Department bruary 2009 05 Fe |
|
|
| Issued b |
|
|
|
|

|
|
| Double click on ‘Device Manager’ |
| yVerified by Approved by |
| FSD R&T Dept OSL Manager |
| Revision 0 |
| 05 Feb 2009 20 Feb 2009 02 Mar 2009 |
| ITEC GEO 0086L14 200T |
|
|
|
|
|
|

|
|
| To see various ports, expand the ‘Ports (COM & LPT)’ |
|
|
|
|
|
|
|
|
|
|

![]()
|
|
| GeoFast FID (Version 2.12) |
| Calibration and Setup Procedure |
| ITEC GEO0086 |
| Page 3/17 |
|
|
| FSD – Field Support Department bruary 2009 05 Fe |
|
|
| Issued b |
|
|

|
|
| Right click on Communication Port (COM1), double click on ‘Properties”. |
|
|
|
|

|
|
| Click on ‘Port settings’ and then double click on ‘Advance’. |
|
|
|
|

|
|
| Set the ‘Transmit Buffer to Low (1). |
| Click OK and again OK and close the Device Manager. Reboot the GFF computer. |
|
|
| Important: The GFF computer COM 1 port is always connected to the GFF gas panel COM port. The GFF |
| computer COM 2 port is connected to the RTM computer (either Persyst board or the COM port) hardware. |
|
|
| yVerified by Approved by |
| FSD R&T Dept OSL Manager |
| Revision 0 |
| 05 Feb 2009 20 Feb 2009 02 Mar 2009 |
| ITEC GEO 0086L14 200T |
|
|

![]()
|
|
| GeoFast FID (Version 2.12) |
| Calibration and Setup Procedure |
| ITEC GEO0086 |
| Page 4/17 |
|
|
| FSD – Field Support Department bruary 2009 05 Fe |
|
|
| Issued b |
|
|
| 2. On the RTM computer – Modification of RTM_RTG.FIC file depending on RTM |
| hardware configuration: |
| 嫈 If Persyst 4 port communication board present, port 1 and port 2 have preset speed of 19200. You can |
| assign any of the two available ports for GFF version 2.12. Example given below. |
|
|
|
|
| RTM_RTG.FIC setting, using a PERSYST board: |
| '----------------------------------------' |
| 'ID Type Port Speed Par Bits Stp ' |
| '----------------------------------------' |
| :PORTS |
| DAP1 DCP88 1 19200 N 8 1 ' Reserved for DAP#1 |
| GASOUT DCP88 2 19200 N 8 1 ' Reserved for GFF windows 2000/XP |
| rem MWD DCP88 3 9600 N 8 1 ' Reserved for MWD |
| rem GASAUX DCP88 4 1200 N 8 1 ' Reserved for Auxillary |
| Rem GASOUT COMX 2 1200 N 8 1 ' Reserved for GasOut |
| rem GASAUX COMX 3 9600 N 8 1 ' Reserved for GasAux (Second Reser |
| rem 7015 COMX 1 9600 N 8 1 ' Reserved for DataScan 7015 |
| rem REMOTE COMX 2 19200 N 8 1 ' Reserved for REMOTE Ascii |
| rem GASIN COMX 2 19200 N 8 1 ' Reserved for GasIn |
| rem ASCII COMX 1 9600 N 8 1 ' Reserved Ascii Port |
| rem WITS0 COMX 2 9600 N 8 1 ' Reserved WITS0 Port |
| rem GDF COMX 1 9600 N 8 1 ' Reserved GDF Display |
| rem CSU COMX 2 1200 N 7 2 ' Reserved Schlumberger |
| rem HPINT2 DCP88 4 1200 N 8 1 ' Reserved for REMOTE Ascii/HPINTEG |
| TCPIP ' Reserved for TCPIP via RTG_NT |
| yVerified by Approved by |
| FSD R&T Dept OSL Manager |
| Revision 0 |
| 05 Feb 2009 20 Feb 2009 02 Mar 2009 |
| ITEC GEO 0086L14 200T |
|
|
|
|
| 嫈 If you are using COM port communication, you can configure any of the ports to 19200 and then assign |
| that port for GFF version 2.12. |
|
|
| RTM_RTG.FIC setting, using a COM ports board: |
| '----------------------------------------' |
| 'ID Type Port Speed Par Bits Stp ' |
| '----------------------------------------' |
| :PORTS |
| rem DAP1 DCP88 1 19200 N 8 1 ' Reserved for DAP#1 |
| rem GASOUT DCP88 2 19200 N 8 1 ' Reserved for DAP#2 |
| rem MWD DCP88 3 9600 N 8 1 ' Reserved for MWD |
| rem GASOUT DCP88 4 1200 N 8 1 ' Reserved for GasOut |
| DAP1 COMX 1 19200 N 8 1 ' Reserved for DAP#1 |
| GASOUT COMX 2 19200 N 8 1 ' Reserved for GFF windows 2000/XP |
| rem GASAUX COMX 3 1200 N 8 1 ' Reserved for Auxillary |
| rem 7015 COMX 1 9600 N 8 1 ' Reserved for DataScan 7015 |
| rem REMOTE COMX 2 19200 N 8 1 ' Reserved for REMOTE Ascii |
| rem GASIN COMX 2 19200 N 8 1 ' Reserved for GasIn |
| rem ASCII COMX 1 9600 N 8 1 ' Reserved Ascii Port |
| rem WITS0 COMX 2 9600 N 8 1 ' Reserved WITS0 Port |
| rem GDF COMX 1 9600 N 8 1 ' Reserved GDF Display |
| rem CSU COMX 2 1200 N 7 2 ' Reserved Schlumberger |
| rem HPINT2 DCP88 4 1200 N 8 1 ' Reserved for REMOTE Ascii/HPINTEG |
| TCPIP ' Reserved for TCPIP via RTG_NT |
|
|
|
|
|
|
|
|

![]()
|
|
| GeoFast FID (Version 2.12) |
| Calibration and Setup Procedure |
| ITEC GEO0086 |
| Page 5/17 |
|
|
| FSD – Field Support Department bruary 2009 05 Fe |
|
|
| Issued b |
|
|
| Main Features of the Calibration Procedures for GFF Version 2.12 |
|
|
| The calibration procedure is split into four sections so as to improve the quality control of gas analysis. |
|
|
| 妈 Prior to calibrating the GFF gas panel |
| In this section the chromatograph cycle is checked for presence of C6. The time when C1 |
| window opens and nC5 window closes are checked if they are in working range. The setting up |
| of each gas component windows, auto tracking and base value are updated on the Setting |
| screen of the GFF. |
| 妈 Chromatograph calibration for GFF |
| In this section the procedure of doing the Chromatograph calibration using the calibration kit is |
| explained in details. |
| 妈 Total Gas multipoint calibration for DRY gas |
|
|
| Here the procedure of using the new multipoint calibration for Total Gas and setting up the Dry |
| gas calibration table are explained. In this dry gas calibration method only Methane gas |
| canister of different concentrations are used with the calibration kit. |
| 妈 Total Gas multipoint calibration for WET gas |
|
|
| In this section the procedure of using the Wet gas multipoint calibration table for Total Gas is |
| explained. In this wet gas calibration method, Mixture gas canisters of different concentration |
| are used with the calibration kit. In this table the physical value for each calibration point is the |
| Methane Equivalent % of that gas canister is entered. |
| Items required in calibrating the GFF gas panel: |
| • Calibration kit. (700.BL.780) |
|
|

|
|
|
|
| • Mixture calibration gas canisters – (700.BV.527) 100ppm Mixture |
| (700.BV.526) 1000ppm Mixture |
| (700.BV.525) 1% Mixture |
| (700.BV.524) 5% Mixture |
| (700.BV.523) 10% Mixture |
|
|
| yVerified by Approved by |
| FSD R&T Dept OSL Manager |
| Revision 0 |
| 05 Feb 2009 20 Feb 2009 02 Mar 2009 |
| ITEC GEO 0086L14 200T |
|
|

![]()
|
|
| GeoFast FID (Version 2.12) |
| Calibration and Setup Procedure |
| ITEC GEO0086 |
| Page 6/17 |
|
|
| FSD – Field Support Department bruary 2009 05 Fe |
|
|
| Issued b |
| • Methane calibration gas canisters – (700.BV.522) 1% Methane |
| (700.BV.521) 10% Methane |
| (700.BZ.939) 30% Methane (optional) |
| (700.BV.520) 100% Methane |
| Note: |
| When you power up the GFF, let it run for few hours so as to have proper temperature stability inside the |
| analysis module. |
| Always do the Chromatograph (CHR) calibration first and then do Total Hydrocarbon Gas (THG) |
| calibration. |
| Check that the GFF air input is set at 3.5 bar pressure. |
| Check that the H2 is set to 0.5 bar pressure. |
|
|
| Things to be checked when you setup a GFF or replace the Analysis module and also when the |
| flows are adjusted. |
|
|
| I. Prior to calibrating the GFF gas panel |
| 1. Check for presence of C6 gas peaks (we do not measure C6 but can detect its peaks when flows are |
| wrongly set). |
| On the “Setting” screen change the cycle time from 42 seconds to 72 seconds. Then click OK. |
|
|

|
|
|
|
| 2. Connect 10% gas mixture canister to the gas line port of the GFF analysis module using the calibration |
| kit regulator and regulate the gas pressure to 1 bar. |
|
|
|
|

|
|

|
|
|
|
|
|
| yVerified by Approved by |
| FSD R&T Dept OSL Manager |
| Revision 0 |
| 05 Feb 2009 20 Feb 2009 02 Mar 2009 |
| ITEC GEO 0086L14 200T |
|
|

![]()
|
|
| GeoFast FID (Version 2.12) |
| Calibration and Setup Procedure |
| ITEC GEO0086 |
| Page 7/17 |
|
|
| FSD – Field Support Department bruary 2009 05 Fe |
|
|
| Issued b |
|
|
| Display the GFF screen on Chromatograph cycle on log scal3. e and check if C6 present. Do this process |
| for 3 complete cycles to confirm there is no presence of C6. |
|
|
|
|

|
|
|
|
| If C6 gas peaks are not present (as above), put back the cycle time to 42 seconds and go to step #5. |
|
|

|
|
|
|
| If the C6 gas peaks are present (as above) than redo the CHR Column Precut Flow again (in this case |
| the Precut flow has to be reduced |
| 4. |
| ). After setting the precut flow, again check the chromatograph cycle |
| for presence of C6 gas peaks. |
|
|
| CHR Column Precut Flow (CCPF) |
|
|
|
|

|
|
| CHR Column Pressure (CCP) |
|
|
|
|
| yVerified by Approved by |
| FSD R&T Dept OSL Manager |
| Revision 0 |
| 05 Feb 2009 20 Feb 2009 02 Mar 2009 |
| ITEC GEO 0086L14 200T |
|
|

![]()
|
|
| GeoFast FID (Version 2.12) |
| Calibration and Setup Procedure |
| ITEC GEO0086 |
| Page 8/17 |
|
|
| FSD – Field Support Department bruary 2009 05 Fe |
|
|
| Issued b |
| 5. Put back the cycle time to 42 seconds. Check the minimum time of C1 gas (open window); the open |
| window of C1 should be between the range of 19 seconds to 22 seconds. If it is not in range then |
| slowly adjust the Column Pressure regulator (clockwise rotation to increase pressure so as to allow the |
| gases to come early in the cycle, counter clockwise for gases to come late). |
|
|

|
|
| For Auto tracking, |
| time in seconds |
| when C1 reach its |
| maximum peak |
| Close nC5 window |
| before 38 secondsMinimum (open) C1 window between 19 |
| to22seconds |
| 6. Check the maximum time (close window) of nC5 gas, this maximum time should be before 38 seconds. |
| If not in range than slowly adjust the Column Pressure regulator. Again check the C1 open window and |
| nC5 close windows, they should both be in the working range. Note that the last 4 seconds of the |
| chromatograph cycle time is required for the GFF computer to do internal processing of data. |
| 7. If iC5 and nC5 gas canister concentration are similar but you observe that the respective curves are of |
| different height then increase the Precut Column Flow by small amount. It is important that you should |
| not allow C6 to be detected. |
|
|

|
|
| iC5 and nC5 gas |
| peaks are of different |
| height |
| 8. |
| Note down the minimum (open) / maximum (close) for C1, C2, C3, iC4, nC4, iC5 and nC5. Also check |
| the C1 peak time for updating the Auto tracking time. Input all this information on the “ |
| Setting” screen |
| table. |
|
|

|
|
| Minus two seconds from |
| the Minimum (open) C1 |
| window value to give you |
| the Base value |
|
|
| yVerified by Approved by |
| FSD R&T Dept OSL Manager |
| Revision 0 |
| 05 Feb 2009 20 Feb 2009 02 Mar 2009 |
| ITEC GEO 0086L14 200T |
|
|

![]()
|
|
| GeoFast FID (Version 2.12) |
| Calibration and Setup Procedure |
| ITEC GEO0086 |
| Page 9/17 |
|
|
| FSD – Field Support Department bruary 2009 05 Fe |
|
|
| Issued b |
|
|
| 9. Now flush / purge the GFF gas panel with compressor air, (use the extension tubing to connect |
| compressor air to the calibration kit regulator) regulate this air pressure to 1 bar in Gas line 1 port of |
| the analysis module. Also flush the Manual Injection port with air. |
|
|
| II. Chromatograph calibration for GFF |
| Since the Calibration port of the GFF analysis module cannot be used for the Chromatograph calibration, we |
| have to use the Manual injection port located in the front of the analysis module. Follow the diagram shown |
| below to line up the gas canister to the analysis module. |
| Manual injection port |
|
|
| yVerified by Approved by |
| FSD R&T Dept OSL Manager |
| Revision 0 |
| 05 Feb 2009 20 Feb 2009 02 Mar 2009 |
| ITEC GEO 0086L14 200T |
|
|
|
|


| Regulate the pressure regulator to set flow of |
| 40cc/min on the Alltech flowmeter. |
|
|

|
|
| ꓸ Connect the Calibration kit tubing to the Manual injection port along with the Alltech flowmeter in series as |
| shown in the above diagram. |
| ꓸ Now connect Air input to the Calibration kit, regulate the pressure regulator so as to have a set flow of 40 |
| cc/min of air (less than 0.5 bar pressure) into the Manual injection port. Since the flow is set do not adjust |
| this pressure regulator while doing the five points chromatograph calibration. |
| ꓸ Connect the required gas canister to calibration kit, flush the kit with little gas and then connect it to the |
| Manual injection port as shown in the diagram below. |
|
|
|
|

|
|

|
|
|
|
| It is a good practice to flush the calibration kit with calibration gas of the next |
| injection so as to get better gas readings. |

![]()
|
|
| GeoFast FID (Version 2.12) |
| Calibration and Setup Procedure |
| ITEC GEO0086 |
| Page 10/17 |
|
|
| FSD – Field Support Department bruary 2009 05 Fe |
|
|
| Issued b |
| On the GFF main screen go to the Hidden Hot Key menu and select the GFF in “Advanced User” mode. Select |
| OK to accept the new setting. |
|
|

|
|

| 1. Go to the Calibration icon so as to activate the Calibration screen as seen below. |
|
|
|
|

| Click Predefined Gas |
| Hidden Hot Key area |
| (between 4 squares) |
| Click Calibration icon |
| 2. Click on ‘Predefined Gas’, click on Injection #1 and input the gas values for C1 to nC5 as labeled on |
| the 100ppm gas canister and click on ‘Update’ to accept the changes for that injection. Do this process |
| for each of the five injections starting from lower range to higher range. (Note: Injection #1 is 100 ppm |
| C1 mixture). |
|
|

| 3. Connect the Calibration kit regulator and the 100 ppm calibration gas canister. Flush the calibration kit |
| tubing with the respective gas and then connect to the Manual injection port of the analysis module and |
| regulate the flow of of 40 cc/min on the Alltech flowmeter as explained earlier. |
|
|
| 4. Click on ‘Run’ for calibration and then click on ‘Yes’ to start the calibration process. |
| yVerified by Approved by |
| FSD R&T Dept OSL Manager |
| Revision 0 |
| 05 Feb 2009 20 Feb 2009 02 Mar 2009 |
| ITEC GEO 0086L14 200T |
|
|

![]()
|
|
| GeoFast FID (Version 2.12) |
| Calibration and Setup Procedure |
| ITEC GEO0086 |
| Page 11/17 |
|
|
| FSD – Field Support Department bruary 2009 05 Fe |
|
|
| Issued b |
|
|


| Each gas component (C1 – nC5) |
| has a grid with black Theoretical |
| curve and respective colour active |
| curve with five injection points |
| Click Run |
| During the ‘Run’ mode, the display on the Gas Composition table starts with 100 ppm C1 as the first |
| predefined gas mixture (injection #1). The gas from the canister flows through the CHR sample loop for |
| @ 10 sec, then the commutation of 6 port valve takes place and analysis process begins for that gas |
| sample. When that cycle completes, the ‘Validate’ box turns from grey colour text to black colour |
| text. Click on ‘Validate’ to accept this calibration point for 100 ppm concentration, see that your |
| calibration points for each of the components are close enough to the black theoretical point. |
| Note: |
| Compare the gas readings displayed on the RTM* computer, these gas readings are base on the |
| last active calibration and not your present calibration which is still not saved. Do not compare the |
| reading for the first two to three cycles, the gas may be contaminated so ignore these cycles. |
|
|
| *use the latest RTM software version. |
|
|
| 5. After validation if you find that this point is not good (as shown in the grid below) then you have to click |
| on ‘Cancel’ before the end of the running cycle to go back to 100 ppm gas composition. Once the point |
| is validated, the gas composition goes to the next predefined 1000 ppm gas mixture. |
|
|

|
|
| 6. Connect the 1000 ppm gas mixture canister to the Calibration kit and flush the calibration kit tubing with |
| gas. Connect the calibration kit to the calibration port and regulate the gas pressure to 1 bar on the |
| pressure gauge. |
| 7. Click on ‘Validate’ to accept this calibration point for 1000 ppm concentration. Follow the same |
| procedure for 1%, 5%, 10% gas mixtures. The purpose of ‘Skip’ is to ignore a particular injection and |
| go to the next injection point. You can skip if you do not have that particular gas mixture canister. |
| 8. After you have validated all the five gas injections then you can click ‘Save’ then click ‘Yes’ to save |
| your new calibration |
|
|
|
|
| yVerified by Approved by |
| FSD R&T Dept OSL Manager |
| Revision 0 |
| 05 Feb 2009 20 Feb 2009 02 Mar 2009 |
| ITEC GEO 0086L14 200T |
|
|

![]()
|
|
| GeoFast FID (Version 2.12) |
| Calibration and Setup Procedure |
| ITEC GEO0086 |
| Page 12/17 |
|
|
| FSD – Field Support Department bruary 2009 05 Fe |
|
|
| Issued b |
|
|
| 9. You can click ‘Stop’ then click ‘Yes’ to stop the calibration process. |
| yVerified by Approved by |
| FSD R&T Dept OSL Manager |
| Revision 0 |
| 05 Feb 2009 20 Feb 2009 02 Mar 2009 |
| ITEC GEO 0086L14 200T |
|
|
|
|
|
|
|
|
| 10. The ‘Restart’ option is to redo all the five point calibration again if needed. |
|
|
|
|




| 11. You can click ‘Advance Calibration’ the file which works in ‘Advance User’ mode. |
|
|
|
|
| 12. Check if all the five calibration points have been saved for C1 – C5’s. |

![]()
|
|
| GeoFast FID (Version 2.12) |
| Calibration and Setup Procedure |
| ITEC GEO0086 |
| Page 13/17 |
|
|
| FSD – Field Support Department bruary 2009 05 Fe |
|
|
| Issued b |
|
|

| All calibration files listed. |
| Note the file with TH is |
| the Theoretical calibration |
| file. |
|
|
| 13. If you intend to use another calibration file then you have to highlight that particular calibration file and |
| click on ‘Select CAL’ button so as to make it active. |
| 14. Add – to make a copy of the highlighted calibration file. Remove – to delete the highlighted file. |
| 15. Update – to save after editing the points (not to be used in the field for modifications) |
| 16. Disconnect the gas canister and flush / purge the GFF gas panel with compressor air, (use the |
| extension tubing to connect compressor air to the calibration kit regulator) regulate this air pressure to |
| 1 bar in Calibration port of the analysis module. |
|
|
|
|
| III. |
| Total Gas multipoint calibration for GFF |
| (Dry gas calibration) |
| 1. Flush / purge the GFF gas panel with compressor air, (use the extension tubing to connect compressor |
| air to the calibration kit regulator) regulate this air pressure to 1 bar in Gas line #1 port of the analysis |
| module. |
|
|

|
|
|
|
| 2. Double click the Total Gas box area to activate the |
| Total Gas calibration window. |
|
|
| yVerified by Approved by |
| FSD R&T Dept OSL Manager |
| Revision 0 |
| 05 Feb 2009 20 Feb 2009 02 Mar 2009 |
| ITEC GEO 0086L14 200T |
|
|

![]()
|
|
| GeoFast FID (Version 2.12) |
| Calibration and Setup Procedure |
| ITEC GEO0086 |
| Page 14/17 |
|
|
| FSD – Field Support Department bruary 2009 05 Fe |
|
|
| Issued b |
|
|

| Click here |
| 3. For Total Gas you have two groups of calibration tables – Dry gas calibration table (using Methane gas |
| canisters only) and Wet gas calibration table (using Mixture gas canisters only). Only one type of |
| calibration table can be kept active. If you need Dry gas then highlight the dry Gas circle and click on |
| ‘Select’ so as to make Dry Gas calibration active. |
|
|

| 4. It is important to purge the GFF gas panel through the gas line #1 port of the Analysis module for few |
| minutes. Observe the Raw μV signal (grey box at the bottom), if this signal is more or less steady |
| (stable) then you click the first ‘UPDATE’ point. This will copy the raw signal value from the bottom to |
| point #1 Raw box. For the Calib box type 0 ppm since air is used to get the Zero - no gas value. After |
| updating the first point of Raw and Calib then click on ‘Validate’ to recognize the new point. Observe |
| the Total Gas display window should show 0 ppm value, if Total Gas display is not showing 0 ppm then |
| redo the raw μV signal value once again to have a proper Zero for the first point. |
|
|

|
|

|
|
|
|
| yVerified by Approved by |
| FSD R&T Dept OSL Manager |
| Revision 0 |
| 05 Feb 2009 20 Feb 2009 02 Mar 2009 |
| ITEC GEO 0086L14 200T |
|
|

![]()
|
|
| GeoFast FID (Version 2.12) |
| Calibration and Setup Procedure |
| ITEC GEO0086 |
| Page 15/17 |
|
|
| FSD – Field Support Department bruary 2009 05 Fe |
|
|
| Issued b |
| 5. Connect 1% Methane gas canister to the Calibration kit. Connect the calibration kit to the Gas line #1 |
| port of the analysis module and regulate the gas pressure to 1 bar on the pressure gauge. |
|
|
|
|

|
|

|
|
| 6. Observe the Raw μV signal (grey box at the bottom), if this signal is more or less steady (stable) then |
| you click the second ‘UPDATE’ point. This will copy the raw signal value from the bottom to point #2 |
| Raw box. For the Calib box type 10000 ppm since the calibration gas canister was 1% Methane. Click |
| on ‘Validate’ to recognize the second point. Observe the Total Gas display window should show @ |
| 10000 ppm value, if not then redo the ‘Update’ for the same concentration of gas and again ‘Validate’. |
|
|

|
|
| 7. Follow the same procedure for 10% and 100% Methane canisters. |
|
|
| IV. |
| Total Gas multipoint calibration for GFF |
| (Wet gas calibration) |
| 1. Flush / purge the GFF gas panel with compressor air, (use the extension tubing to connect compressor |
| air to the calibration kit regulator) regulate this air pressure to 1 bar in Gasoline #1 port of the analysis |
| module. |
|
|

|
|
| yVerified by Approved by |
| FSD R&T Dept OSL Manager |
| Revision 0 |
| 05 Feb 2009 20 Feb 2009 02 Mar 2009 |
| ITEC GEO 0086L14 200T |
|
|

![]()
|
|
| GeoFast FID (Version 2.12) |
| Calibration and Setup Procedure |
| ITEC GEO0086 |
| Page 16/17 |
|
|
| FSD – Field Support Department bruary 2009 05 Fe |
|
|
| Issued b |
| 2. Click the Total Gas box area to activate the |
| Total Gas calibration window. |
|
|
|
|

| Click here |
| 3. If you need Wet gas then highlight the wet Gas circle and click on ‘Select’ so as to make Wet Gas |
| calibration active. |
|
|

|
|
|
|
| 4. It is important to purge the GFF gas panel through the gas line #1 port of the Analysis module for few |
| minutes. Observe the Raw μV signal (grey box at the bottom), if this signal is more or less steady |
| (stable) then you click the first ‘UPDATE’ point. This will copy the raw signal value from the bottom to |
| point #1 Raw box. For the Calib box type 0 ppm since air is used to get the Zero - no gas value. After |
| updating the first point of Raw and Calib then click on ‘Validate’ to recognize the new point. Observe |
| the Total Gas display window should show 0 ppm value, if Total Gas display is not showing 0 ppm then |
| redo the raw μV signal value once again to have a proper Zero for the first point. |
|
|

|
|

|
|
|
|
| yVerified by Approved by |
| FSD R&T Dept OSL Manager |
| Revision 0 |
| 05 Feb 2009 20 Feb 2009 02 Mar 2009 |
| ITEC GEO 0086L14 200T |
|
|

![]()
|
|
| GeoFast FID (Version 2.12) |
| Calibration and Setup Procedure |
| ITEC GEO0086 |
| Page 17/17 |
|
|
| FSD – Field Support Department bruary 2009 05 Fe |
|
|
| Issued b |
| 5. Connect 1000 ppm Mixture gas canister to the Calibration kit. Connect the calibration kit to the |
| Gasoline #1 port of the analysis module and regulate the gas pressure to 1 bar on the pressure gauge. |
|
|
|
|

|
|

|
|
| 6. Observe the Raw μV signal (grey box at the bottom), if this signal is more or less steady (stable) then |
| you click the second ‘UPDATE’ point. This will copy the raw signal value from the bottom to point #2 |
| Raw box. Since 1000 ppm mixture gas canister will give 6750 ppm Methane Equivalent % (C1 + 2xC2 |
| + 3xC3 + 4xiC4 + 4xnC4 + 5xiC5 + 5xnC5). In the Calib box you have to input Methane Equivalent |
| value. Click on ‘Validate’ to recognize the second point. Observe the Total Gas display window should |
| show @ 6750 ppm value, if not then redo the ‘Update’ for the same concentration of gas and again |
| ‘Validate’. |
|
|
|
|

|
|
| Calculate the Methane Equivalent |
| % base on the canister values for |
| each component. |
|
|
| 7. Follow the same procedure when calibrating the 1%, 5% and 10% Mixture gas canisters. |
|
|
| Mixture gas canister Methane Equivalent % |
| 1000 ppm C1 Mixture 6750 ppm |
| 10000 ppm C1 Mixture 67500 ppm |
| 50000 ppm C1 Mixture 200500 ppm |
| 100000 ppm C1 Mixture 330000 ppm |
| Note: Since the 10% Mixture has the C6 components, it has to be included in the Methane equivalent %. |
| 8. Remove the calibration kit from gasline #1 port of the analysis module and connect back the gasline #1 |
| tubing coming from the pump module. Start both the pumps and flush out all the contaminated gas |
| from the system. Now the GFF calibration is complete. |
|
|
| yVerified by Approved by |
| FSD R&T Dept OSL Manager |
| Revision 0 |
| 05 Feb 2009 20 Feb 2009 02 Mar 2009 |
| ITEC GEO 0086L14 200T |
|
|
No hay comentarios:
Publicar un comentario